Kasba, Kolkata, West Bengal
Leading Manufacturers, Exporters and Wholesaler of Five Tube Flowmeter, Flowmeter Tubes, Single Tube Flowmeter, Dual Tube Flowmeter, Dual Tube Flowmeter with AHD, Dual Tube Cascade Flowmeter, Four Tube Flowmeter, Variable Area Flowmeter, Anaesthesia Rotameters Units and Anaesthetic Rotameters Units With Anti Hypoxia from Kolkata.
Business Type | Manufacturer, Exporter, Supplier |
Oxygen Inlet | 6 mm Festo connection from rear bottom ,(Straight) |
N2o Inlet | 6 Mm Festo Connection from Rear Bottom ,(straight) |
Air Inlet | 6 Mm Festo Connection from Rear Bottom ,(straight) |
Mixed Gas Outlet | 6 Mm Festo Connection from Rear ,Top,(straight) |
Control Knob Oxygen | Aluminium with Special Profile as per Iso 5358, |
Control Knob N2O | Aluminium Serrated, Serration Depth <1mm, Knob Dia Lower Than Oxygen Knob Dia |
Control Knob Air | Aluminium, serrated, serration depth <1mm, knob dia lower then Oxygen knob Dia. |
Working pressure (max) | 420 kPa (N2O, O2 &AIR) |
Oxygen Tube Range Low Flow | 0.1 – 1 L/min, with Graduations in Lpm Unit |
Oxygen Tube Range High Flow | 1 – 10 L/min, with Graduations in Lpm Unit |
N2o Tube Range Low Flow | 0.1 – 1 L/min, with Graduations in Lpm Unit |
N2o Tube Range High Flow | 1 – 12 L/min, with Graduations in Lpm Unit |
Air Tube Range of Flow | 0.1 – 15 L/min Graduations in Lpm Unit |
Metering Tube | Borosilicate Glass. |
Float | Aluminium-is-6063,Anodised Red –for O2, N2o & Air |
Accuracy (o2, N2o & Air) | As per Iso-5358 |
Tube Length (o2, N2o & Air) | 220 Mm |
Markings/identification | Float Mark On Tubes & other Markings as per Iso 5358 |
Mounting | Panel Mounting, as per Drawing |
Housing | Aluminum Frame, Acrylic Front, Background On White Acrylic Sheet. |
Material of Construction | Brass-is-319 Natural Colour/aluminium Is-6063 Anodized Natural, O Rings & Washer Epdm |
Tube Fitting Arrangement | Left-o2, Middle-n2o, Right –air (o2.lf-o2.hf) –(n2o.lf-n2o.hf)- (air) |
Hypoxia Unit | Geared Type Fitted with the Frame, |
N2o Cutoff | Built-in Fitted At Back |
Oxygen / N2o Ratio | 30 + 5% of O2 At all Settings After N2o Flow Opening. |
Labels for Control Knob-o2/n2o/air | O2-black On White Back Ground/n2o-white On Blue Back Ground /air-white On Black Back Ground. |
Compliance To | Iso-5358, Rohs |
Download | View |
Download 01 | View |
Preferred Buyer From
Location | Worldwide |
Anaesthetic flowmeters are available in 2 – 5 tubes versions. The flowmeters (2-5 tube versions) can be supplied with or without Anti Hypoxia Device.
The housings are made from Aluminum and fitted with fine adjustment valves for easy control of the anaesthetic gases. The front cover is made from moulded acrylic. The devices have a phosphorescent “glow in the dark” (normally white) background for easy reading even in case of sudden darkness.
The knobs are made from finest quality ABS. The Oxygen knob has a touch coded profile and all the knobs are colour coded as per ISO 5358 requirements.
Internal gas connections are made using polyurethane tubes and Festo type connections. A non return valve in the downstream of the AHD ensures no flow of N2O in the respiratory circuit in the event of single fault condition when Oxygen line gets blocked.
Various options of inlet and outlet connections are available
Business Type | Manufacturer, Exporter, Supplier |
Calibrated for | L/min at 20 Deg C / 1013 mbar abs |
Conform to | ISO 5358 |
Bobbin Float | Aluminium Alloy Anodised Red, with Silver Dot for Easily Observing Rotation |
Preferred Buyer From
Location | Worldwide |
Optimized & Industry-ready Solutions
Flowmeter are used for measuring flow in Anesthesia Machine
Specification:
Business Type | Manufacturer, Exporter, Supplier |
Inlet | Nipple-6 Mm |
Outlet | 23mm Tapper Outlet |
Control Knob | Aluminium Anodized , with Special Profile as per Our Drg No.kn-016, Iso-5358 |
Mounting | Panel |
Trim Material | Aluminium Anodized , Brass Nickel Plated, Front Clear & Back White Acrylic |
Metering Tube | Borosilicate Glass Antistatic Coated.(resistance Within < 5 M) |
Float | Aluminium Anodized Red |
Packing | Neoprene Rubber 50-60 Hs |
Flow Range | 0.1-10 LPM O2 (Dual range) |
Operating Temp | 20 Deg C |
Working Pressure | 4 .2 Kg/cm2g |
Accuracy | + 10 % of Indicated Value or 300 Ml/min Whichever is Higher (iso-5358) |
Calibration | Calibrated at STP (20 Deg C&1013mb abs) |
Measurement | As per Drg: MO-1300 |
Compliance | Iso-5358/ Rohs |
Preferred Buyer From
Location | Worldwide |
Anaesthetic flowmeters are available in single tubes versions. Available in various model.
Anaesthetic Rotameters are used for measuring the rate of flow of gases used for anaesthesia. The gases measured are Oxygen and Air.
Business Type | Manufacturer, Exporter, Supplier |
Accuracy | Compliant to ISO 5358, or ISO 15002, or VDI 3513 (upto G=2.5, lin=50%) |
No. of Tubes | 1-tube to 6-tubes available as standard |
Tube length | Borosilicate glass, Standard 230mm. Other sizes on request (See tube catalogue) |
Antistatic | Conductive tubes and bushes for static dissipation |
Float | Anodised Aluminum, Red, rotating as standard. |
Anti Hypoxia | 2-tube to 6-tube versions are also available with Antihypoxia device. |
FA valve | Available with Fine adjustment valves for accurate flow control |
Gases | Oxygen, Nitrous Oxide, Air, Xenon |
Background | Yellow green “Glow in the dark” for easy viewing |
Body | Anodised aluminum, or powder coated, as per your choice. |
Front | Front cover made from transparent acrylic sheet |
Knob | Color and touch coded as per ISO 5358 |
Download | View |
Preferred Buyer From
Location | Worldwide |
Anaesthetic Rotameters are used for measuring the rate of flow of gases used for anaesthesia. The gases measured are Oxygen, Nitrous oxide, Carbon dioxide, Xenon, and Air.
Anaesthetic Rotameters units with control valve system can be of 1 tube, 2 tubes, 3 tubes and so on as per user requirement. It comes as a complete unit for fitting on the Anaesthesia machine.
Flow Range
Connections
Business Type | Manufacturer, Exporter, Supplier |
Oxygen Inlet | 6 Mm Festo Connection from Rear Bottom ,(straight) |
N2o Inlet | 6 Mm Festo Connection from Rear Bottom ,(straight) |
Mixed Gas Outlet | 6 Mm Festo Connection from Rear Top ,(straight) |
Control Knob Oxygen | Aluminium with Special Profile as per Iso 5358 |
Control Knob N2o | Aluminium Serrated, Serration Depth <1mm, Knob Dia Lower Than Oxygen Knob Dia |
Working Pressure (max) | 420 Kpa (both N2o & O2) |
Oxygen Tube Range | 0.1 – 10 L/min, with Expanded Scale At 0.1 – 1 L/min Graduations in Lpm Unit |
N2o Tube Range | 0.1– 12 L/min, with Expanded Scale At 0.1 – 1 L/min Graduations in Lpm Unit |
Metering Tube | Borosilicate Glass. |
Float | Aluminium-is-6063,Anodised Red –for O2& N2o |
Accuracy (o2& N2o) | As per Iso-5358 |
Tube Length (o2& N2o) | 220 Mm |
Markings/identification | Float Mark On Tubes & other Markings as per Iso 5358 |
Mounting | Back Bar Mounting, as per Drawing |
Housing | Aluminum Frame, Acrylic Front, Background On White Acrylic Sheet. |
Material of Construction | Brass-is-319 Natural Colour/aluminium Is-6063 Anodized Natural, O Rings & Washer Epdm. |
Tube Fitting Arrangement | Left-n2o, Right –o2 |
Hypoxia Unit | Geared Type Fitted with the Frame. |
N2o Cutoff | Built-in Fitted At Back |
Oxygen / N2o Ratio | 30 + 5% of O2 At all Settings After N2o Flow Opening. |
Labels for Control Knob-o2/n2o | O2-black On White Back Ground/n2o-white On Blue Back Ground |
Compliance To | Iso-5358, Rohs |
Download | View |
Preferred Buyer From
Location | Worldwide |
Function
The Flow Head incorporates the following Functions
GENERAL OPERATION
The Flow Head integrates O2 and N20.
The 02 Air and N20 needle valves gas supply pressure is externally regulated to 400Kpai10Kpa as per standard.
The N20 channel incorporates a pressure cut-oft valve. The 02 supply pressure in the circuit controls the output pressure of the pressure cut-off valve. Under all conditions of operation, the N20 flow shall reach zero before the 02 flow. In this flow head the N20 pressure switch shall cut off the flow of N20 before the 02 pressure drops below 50:15Kpa.
PERFORMANCE STANDARDS
Standard flow conditions defined at 20°C, 101 .3Kpa, (760mmHg) Conforms to ISO 5358 requirements.
Flow Tubes
Flow Control Valves
PRESSURE CUT-OFF VALVE
ASSEMBLY
Leakage from atmosphere shall not exceed 108ml/min. when a vacuum of 80cm H20 is applied to all internal.
MATERIALS
Business Type | Manufacturer, Exporter, Supplier |
Inlet | Festo-6mm |
Outlet | 23 Mm Tapper Nozzle |
Control Knob | Aluminium with Special Profile as per Iso 5358, |
Mounting | Panel Mounting |
Trim Material | Brass-is-319 Natural Colour/aluminium Is-6063 Anodized Natural, O Rings & Washer Epdm |
Metering Tube | Borosilicate Glass |
Metering Tube Size | Low Flow-10x230, High Flow-15x230 |
Float | Aluminium Anodized Red |
Packing | Epdm- 50-60 Hs |
Flow Range Low Flow | 0.1-1 Lpm O2 with Graduation 0.05 Lpm |
Flow Range High Flow | 1-10 Lpm O2 with Graduation 0.5 Lpm |
Operating Temp | 10-50 Deg C |
Working Pressure | 4 Kg/cm2g |
Accuracy | + 10 % of Indicated Value or 300 Ml/min Whichever is Higher (iso-5358) |
Calibration | Calibrated At Stp (20 Deg C&1013mb Abs) |
Measurement | As per Drg: M-2501-35x90x282 mm |
Compliance | Iso-5358/ Rohs |
Download | View |
Preferred Buyer From
Location | Worldwide |
Anaesthetic Rotameters are used for measuring the rate of flow of gases used for anaesthesia. The gases measured are Oxygen and Air.
Anaesthetic Rotameters units with single control valve system of 2 tubes as per user requirement. It comes as a complete unit for fitting on the Anaesthesia machine for veterinary used .
1, 1.2, 1.5 & 2 | Nitrous Oxide Gas Max. Flow LPM |
10,12,15& 20 | Total Flow Below 1 L/min. |
Min. 21% | Total Flow Below 1 L/min. & Above |
Download | View |
Preferred Buyer From
Location | Worldwide |
FUNCTION
GENERAL OPERATION
The Flow Head integrates 02 and N20 (the N20 is optional).
The O2 Air and N20 needle valves gas supply pressure is externally regulated to 400Kpa¢10Kpa as per standard.
The N20 channel incorporates a pressure cut-off valve. The O2 supply pressure in the circuit controls the output pressure of the pressure out-oft valve. Under all conditions of operation, the N20 flow shall reach zero before the O2 flow. ln this flow head the N20 pressure switch shall cut oft the flow of N20 before the O2 pressure drops below 50115Kpa.
Business Type | Manufacturer, Exporter, Supplier |
calibrated for | L/min At 20 Deg C / 1013 Mbar Abs |
Conform | ISO 5358 |
Bobbin Float | Special Design Float for Rapid Response to Flow Changes |
Printing | Machine Manufacturer Logo Printed if Required |
Gas Coding | Colour Coding for Each Gas |
Anti Static | Available |
Download 01 | View |
Preferred Buyer From
Location | Worldwide |
Flowmeter are used for measuring flow in Anesthesia Machine
Featurs:
Business Type | Manufacturer, Exporter, Supplier |
Type | Anaesthesia Machine |
Accuracy | Compliant to ISO 5358, or ISO 15002, or VDI 3513 (upto G=2.5, lin=50%) |
No. of Tubes | 1-tube to 6-tubes available as standard |
Tube length | Borosilicate glass, Standard 230mm. Other sizes on request (See tube catalogue) |
Antistatic | Conductive tubes and bushes for static dissipation |
Float | Anodised Aluminum, Red, rotating as standard. |
Anti Hypoxia | 2-tube to 6-tube versions are also available with Antihypoxia device. |
FA valve | Available with Fine adjustment valves for accurate flow control |
Flow Range | High flow tubes are available in ranges of 0-10, 0-12. 0-15 LPM Low flow tubes are available in rang |
Gases | Oxygen, Nitrous Oxide, Air, Xenon |
Connections | Inlet and outlet connections are available as per your choice. Most common are ISO taper with RH ou |
Background | Yellow green “Glow in the dark” for easy viewing |
Body | Anodised aluminum, or powder coated, as per your choice. |
Front | Front cover made from transparent acrylic sheet |
Description | View |
Download 01 | View |
Download 02 | View |
Preferred Buyer From
Location | Worldwide |
Description:
Anaesthetic Rotameters are used for measuring the rate of flow of gases used for anaesthesia. The gases measured are Oxygen, Nitrous oxide, Carbon dioxide, Xenon, and Air
Anaesthetic Rotameters units with control valve system can be of 1 tube, 2 tubes, 3 tubes and so on as per user requirement. It comes as a complete unit for fitting on the Anaesthesia machine.
Business Type | Manufacturer, Exporter, Supplier |
Oxygen Inlet | 6 mm Festo connection from rear bottom ,(Straight) |
N2O Inlet | 4 mm Festo connection from rear bottom ,(Straight) |
Air Inlet | 6 mm Festo connection from rear bottom ,(Straight) |
Mixed gas outlet | 10 mm Festo connection from rear ,Top right,(Straight) |
Control Knob Oxygen | Aluminium with special profile as per ISO 5358, |
Control Knob N2O | Aluminium serrated, serration depth <1mm, knob dia lower than Oxygen knob dia |
Control Knob Air | Aluminium, serrated, serration depth <1mm, knob dia lower then Oxygen knob Dia. |
Working pressure (max) | 420 kPa (N2O, O2 &AIR) |
Oxygen Tube Range Low Flow | 0.1 – 1 l/min, with Graduations in lpm unit |
Oxygen Tube Range High Flow | 1 – 10 l/min, with Graduations in lpm unit |
N2O Tube Range Low Flow | 0.1 – 1 l/min, with Graduations in lpm unit |
N2O Tube Range High Flow | 1 – 12 l/min, with Graduations in lpm unit |
Air Tube Range of Flow | 0.1 – 15 l/min Graduations in lpm unit |
Metering Tube | Borosilicate glass-antistatic coated. |
Float | Aluminium-IS-6063,Anodised red –for O2, N2O & AIR |
Accuracy (O2, N2O & AIR) | As Per ISO-5358 |
Tube length (O2, N2O & AIR) | 230 mm |
Markings/identification | Float mark on tubes & other markings as per ISO 5358 |
Mounting | Back bar mounting, as per drawing |
Housing | Aluminum frame with 3 side acrylic front, Luminous background on white acrylic sheet, printed as per |
Material of construction | Brass-IS-319 nickel plated/aluminium IS-6063 anodized black, back plate Powder coated (black), Ratio |
Tube fitting arrangement | Left-N2O, Middle- Air, Right –O2 (N2O.LF-N2O.HF) - (AIR)-(O2.HF-O2.LF) |
Hypoxia Unit | Pneumatic Type Fitted with the Frame, Diaphragm -Neoprene: Molded rubber as per Drg: DPD-004 |
N2O flow opening/closing | Opening at 0.1-0.4 l/min Oxygen flow, while increasing Closing at 0.1-0.3 l/min Oxygen flow, while d |
Oxygen / N2O ratio | 30 + 10% of O2 at all settings after N2O flow opening. N2O Max Flow 12 LPM |
Labels for Control Knob-O2/N2O/Air | O2-Black on white back ground/N2o-white on Blue back ground /Air-white on black back ground. |
Compliance to | RoHS-3 |
Download | View |